Common Problem | Question Type | Possible Cause | Solution |
FAQ for SpO2 Sensor &Extension Cable | I,No Blood Oxygen,No luminescence | 1.Luminescent pipe was broken | Contact manufacturer |
2.The luminous tube side cable short circuit | Contact manufacturer |
3.Machine-end connection is bad | Connect the ends of the machine |
4.Incorrect polarity | Contact manufacturer |
5.The plug is too loose, could not receive the host signal | Contact manufacturer |
II,No Blood Oxygen and waveform | 1.Receiver tube anomalies | Contact manufacturer to repair |
2.Unstable, luminescence, receiver tube circuit breaker | Contact manufacturer to repair |
3.Incompatible | Choose the right compatible model |
4.Light-Leaking Seriously | Re-use according to the instruction manual |
5.Probe selection error | According to the machine instructions to select the correct probe |
6.Testing position there is no alignment LEDs and receiver | Re-use according to the instruction manual |
7.Patient Attention is too low try several measurements | Please go to the hospital for accurate diagnosis |
8.The fingers are trembling or the patient's body is moving | Don’t Move |
III,The skin of the test site is red and swollen | The probe has not been loosened for a long time | Long term test finger where you need change the test every 4 hours |
IV,display on Blood oxygen values,luminescence,but the blood oxygen is low | Incorrect polarity | Contact manufacturer |
FAQ for ECG Cable and lead wires | I,ECG cannot be displayed correctly | 1.Lead wires or ECG Cable cannot compatible | Please contact our sales to select correct model |
2.Cable doesn’t work,incorrect polarity | Contact supplier |
3.Machine end or patient end connect incorrectly | To connect the ECG cable well |
II,The waveform is unstable and the burr is too more | 1.Some other machine interfere around | Remove interfering machine |
2.Patient is shaking | Make patient stop to shake and has remained relatively static |
3.Cable anti-interference is poor | To contact supplier |
NIBP CUFF & see Attachment for common faults | The blood pressure showed incorrect | 1.The machine end connection incorrect | Connect the ends of the machine |
2.Patient is shaking | Patient stop to shake and has remained relatively static |
3.Repeated test in a short time | Re-test after 15minutes |
4.Host exception | Replace the host |
5.Testing parts of the sleeves tied up is incorrect | According to the icon shows and binding the position of cuff |
6.Selected a wrong size of cuff | To select a suitable cuff according to patient and machine |
FAQ for Temp Probe and extension Cable | Temperature shows incorrect | 1.Host exception | Replace the host |
2.Probe or connect cable incompatible with machine | Select compatible model according to manual illustration |
3.Probe or cable has broken | To replace new probe or extension cable |
4.Machine end connect incorrect | Connecting the cable and end of machine well |